The Titan-OL (On-Line) system provides real-time extractive-gas-phase analysis for online and continuous multi-point analysis. It includes a rugged, industrial Fourier transform infrared (FTIR) spectrometer, gas cell(s), sample manifold, valve-control electronics, and software. The hardware is encased in a NEMA-rated enclosure or in a standard industrial rack-mount cabinet. The manufacturer’s exclusive fiber-optic distribution electronics provide a single hub for temperature, pressure, mass flow, external alarm, and optical switching signals. The system can accommodate up to four 12-port manifolds for a total of 48 sample, purge and calibration lines, with communication between the hub and control computer via fast, stable fiber-optic cable. Titan-OL systems use dual analog-to-digital converter (ADC) architecture and advanced mirror drive control for increased sensitivity and very low detection levels for a variety of compounds. Custom gas cells are corrosion-resistant and are available with path lengths from 0.01 m to 30 m. The systems can operate at temperatures from ambient to 240ºC, allowing moist sample streams to be analyzed without condensation. The included AutoQuant Pro software package handles method calibration, data collection, trend analysis charting, manifold control, and integration of alarm and control functions with plant process hardware and software
Titan-OL (On-Line) system
MIDAC Corp.